Detector Type: Heated
Reaction Chamber, Cooled Photomultiplier Tube Sensor
Gas
Sample System: Stainless
Steel, 191 degree C filter and flow control components
Reaction
Chamber: 65
degree C Stainless Steel reaction chamber
Ranges:
0-4,
10, 40, 100, 400, 1000, 4000, 10000 ppm NO / NOx, Auto-ranging
standard
Response
Time: 1.5s
(T5 to T95)
Gas
Flow Rate: 0.5
to 3.0 lpm
Inlet
Pressure: (-5.0)
to 7.0 psig, calibration gas: -5 to 15 psig
Accuracy
and Repeatability: +/-
1% of range or 0.2ppm
Temperature
Effects: Zero:
<+/- 0.5% from 5 to 35 degrees C, Span: <+/- 0.33% per degreeC
Detector
Noise: <+/-
0.5ppm or 0.1% of range
Linearity:
<
2.0% of point or 1.0% FS
NOX
Gas Converter Efficiency: Greater
than 95%, test with
NOXGEN tester
Remote
Control: Range,
calibration and standby mode via contact closure or RS-232
Outputs:
0-10VDC
and isolated 4-20mA concentration and range ID voltage level
Logic
Outputs: Range,
calibration, gas path, mode and readiness
Alarms:
Two
configurable concentration alarms, logic outputs and one fault contact
set
Power:
Dual
voltage 115/230 VAC +/- 15%, switch-selectable, 600VA max.
Environment:
Temperature:
5-40 degrees C, 95%RH, non-condensing
Dim:
19"
rack-mount, 3RU (133.5mm) height, 525mm depth, 35Kg weight
Options:
NOXGEN Tester,
NO/NO2/NOx Fast Switching Measurement
Warranty:
2
Years Parts, labor and complimentary maintenance service